FDA Adverse Event Summary report: N

INGENUITY CT

MDR report key: 3132767 · Received May 24, 2013

Report

Report Number
1525965-2013-00137
Date Received
May 24, 2013
Report Date
April 29, 2013
Manufacturer
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
Product Code
JAK
PMA / PMN Number
K033326
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NOTE: WE HAVE NOT COMPLETED OUR INVESTIGATION OF THIS EVENT. WE WILL FILE A FOLLOW-UP MDR AT THE COMPLETION OF THE INVESTIGATION. (B)(4).

Description of Event or Problem · 1

PHILIPS RECEIVED INFORMATION FROM THE FIELD SERVICE ENGINEER (FSE) WHO WAS PERFORMING SERVICE ON THE PHILIPS CT SYSTEM WITH THE FRONT GULL-WING COVER OPEN TO ACCESS THE ROTATING FRAME. THE SYSTEM IS DESIGNED TO BE ABLE TO OVERRIDE THE GANTRY FRONT COVER GULL-WING SWITCH BY "PULLING IT OUT". THE FSE PULLED THE SWITCH OUT TO ENERGIZE THE GANTRY. THE GULL-WING SAFETY "FRIDGE" SWITCH SLIPPED OUT OF THE METAL ENCLOSURE AND THE SWITCH CONTACT SHORTED TO THE CHASSIS (THERE IS 120VAC ON SWITCH CONTACTS). THIS SHORT CAUSED AN OPEN FOIL RUN ON THE AC CONTROL UNIT. THE FSE CONFIRMED HE WAS NOT HARMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
231599 INGENUITY CT JAK PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. 728306

Patients

Seq Age Sex Outcome Treatment
1