ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2013-07083
- Event Type
- Malfunction
- Date Received
- May 29, 2013
- Report Date
- May 3, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
SUBMITTED: (B)(4) 2013. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. DEVICE EVALUATION: ON EXAMINATION, THE KEYPAD WAS INTACT. ON TESTING, ALL THE KEYPAD BUTTONS HAD NORMAL RESPONSE. THE KEYPAD COVER WAS REMOVED FOR INVESTIGATION AND REVEALED CONTAMINATION UNDER THE CONTACTS OF ALL THE KEYPAD BUTTONS. THE DISPLAY WAS NOTED TO BE DIM, FADED AND HAVE PINK CONTRAST. A TEST DISPLAY WAS ATTACHED TO PUMP AND ILLUMINATED PROPERLY WITHOUT DISCOLORATION.
THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(6) 2013: THE DISPLAY WAS DIM, FADED AND HAD PINK COLOR AND CONTAMINATION WAS FOUND UNDER THE CONTACTS OF ALL THE KEYPAD BUTTONS.