FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 3132734 · Received May 29, 2013

Report

Report Number
2531779-2013-07083
Event Type
Malfunction
Date Received
May 29, 2013
Report Date
May 3, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

SUBMITTED: (B)(4) 2013. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. DEVICE EVALUATION: ON EXAMINATION, THE KEYPAD WAS INTACT. ON TESTING, ALL THE KEYPAD BUTTONS HAD NORMAL RESPONSE. THE KEYPAD COVER WAS REMOVED FOR INVESTIGATION AND REVEALED CONTAMINATION UNDER THE CONTACTS OF ALL THE KEYPAD BUTTONS. THE DISPLAY WAS NOTED TO BE DIM, FADED AND HAVE PINK CONTRAST. A TEST DISPLAY WAS ATTACHED TO PUMP AND ILLUMINATED PROPERLY WITHOUT DISCOLORATION.

Description of Event or Problem · 1

THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(6) 2013: THE DISPLAY WAS DIM, FADED AND HAD PINK COLOR AND CONTAMINATION WAS FOUND UNDER THE CONTACTS OF ALL THE KEYPAD BUTTONS.