FDA Adverse Event Malfunction Summary report: N

ORTHO SUMMIT SAMPLE HANDLING SYSTEM

MDR report key: 3132629 · Received May 29, 2013

Report

Report Number
2250051-2013-00150
Event Type
Malfunction
Date Received
May 29, 2013
Date of Event
April 29, 2013
Report Date
May 28, 2013
Manufacturer
HAMILTON BONADUZ AG
Product Code
MMH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON (B)(4) 2013 AN OCD FIELD ENGINEER (FE) ARRIVED AT THE CUSTOMER SITE AND FOUND THE TIP SLEEVE STICKING AT POSITION #1. THE FE REPLACED THE TIP SLEEVE AT POSITION #1. THE FE RAN AND PASSED THE SPLLD AND USER SOFTWARE WITH OUT ERRORS. REPAIRS HAVE RETURNED THIS INSTRUMENT TO EXPECTED OPERATION.

Description of Event or Problem · 1

THE ORTHO SUMMIT SAMPLE HANDLING SYSTEM INSTRUMENT DID NOT PIPETTE THE CORRECT AMOUNT OF SAMPLE AND DID NOT GENERATE AN ERROR MESSAGE. NO DEATH OR SERIOUS INJURY WAS ASSOCIATED WITH THIS INCIDENT. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
235648 ORTHO SUMMIT SAMPLE HANDLING SYSTEM SAMPLE PIPETTOR MMH HAMILTON BONADUZ AG 936400