FDA Adverse Event Malfunction Summary report: N

SMALL BATTERY DRIVE

MDR report key: 3132592 · Received May 29, 2013

Report

Report Number
8030965-2013-02416
Event Type
Malfunction
Date Received
May 29, 2013
Date of Event
March 1, 2013
Report Date
March 6, 2013
Manufacturer
SYNTHES GMBH
Product Code
HWE
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: A SERVICE HISTORY OF THE PAST SIX MONTHS HAS BEEN REVIEWED. NO SERVICE HISTORY REVIEW CAN BE PERFORMED. THE ITEM HAS NOT PREVIOUSLY BEEN SENT IN FOR SERVICE WITHIN THE LAST SIX MONTHS. THERE IS NO INFORMATION RELEVANT TO THE CURRENT COMPLAINED ISSUE. (B)(6). PLACEHOLDER.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. ADDITIONAL NARRATIVE: THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. THE MANUFACTURING DOCUMENTS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. DEVICE WAS RECEIVED FOR EVALUATION. INVESTIGATION COORDINATED BY SYNTHES (B)(4). REPORT RECEIVED INDICATES THE REPORTERS COMPLAINT OF A STICKY TRIGGER WAS CONFIRMED. THE SMALL BATTERY DRIVE DRILL WAS TESTED AND THE COMPLAINT WAS DUPLICATED. EVIDENCE INDICATES THIS IS DUE TO USAGE WEAR OVER TIME. PLACEHOLDER.

Description of Event or Problem · 1

IT WAS REPORTED DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION PROCEDURE ON (B)(6) 2013; THE TRIGGER OF THE SMALL BATTERY DRIVE WAS STICKING AND THE DEVICE WAS CONTINUOUSLY RUNNING. IT WAS ALSO REPORTED THE EVENT CAUSED A DELAY IN PROCEDURE BY APPROXIMATELY FIVE MINUTES WHILE THE SURGEON SWITCHED TO ANOTHER DEVICE. THE PROCEDURE WAS COMPLETED WITH NO ADVERSE EVENT TO PATIENT. THIS REPORT IS FOR A SMALL BATTERY DRIVE. THIS IS 1 OF 1 DEVICE FOR THIS EVENT REPORTED ON COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
236697 SMALL BATTERY DRIVE HWE SYNTHES GMBH 001971

Patients

Seq Age Sex Outcome Treatment
1 21 YR