FDA Adverse Event
Malfunction
Summary report: N
CORE UNIVERSAL DRIVER
MDR report key: 3132582
·
Received May 29, 2013
Report
- Report Number
- 0001811755-2013-01218
- Event Type
- Malfunction
- Date Received
- May 29, 2013
- Date of Event
- May 15, 2013
- Report Date
- May 15, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
UPON DISASSEMBLY FOR VISUAL INSPECTION CORROSION WAS FOUND ON THE POTTED TRIGGER ASSEMBLY.
Additional Manufacturer Narrative · 1
INVESTIGATION IS IN PROGRESS. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION HAS BEEN COMPLETED. INVESTIGATION IS IN PROGRESS. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION HAS BEEN COMPLETED.
Description of Event or Problem · 1
WHILE TESTING THE CORE UNIVERSAL DRIVER IT CONTINUED TO RUN WHEN THE TRIGGER WAS NO LONGER ACTIVATED. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.
Description of Event or Problem · 1
WHILE TESTING THE CORE UNIVERSAL DRIVER IT CONTINUED TO RUN WHEN THE TRIGGER WAS NO LONGER ACTIVATED. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 234910 | CORE UNIVERSAL DRIVER | DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE | ERL | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |