FDA Adverse Event Malfunction Summary report: N

CORE UNIVERSAL DRIVER

MDR report key: 3132547 · Received May 29, 2013

Report

Report Number
0001811755-2013-01214
Event Type
Malfunction
Date Received
May 29, 2013
Date of Event
May 14, 2013
Report Date
May 14, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON DISASSEMBLY FOR VISUAL INSPECTION THE BEARINGS WERE WORN AND THE REAR MOTOR ASSEMBLY NEEDED TO BE REPLACED.

Description of Event or Problem · 1

THE CORE UNIVERSAL DRIVER WAS RETURNED FOR SERVICE. UPON EVALUATION AT THE MANUFACTURER IT DISPLAYED A BIAS CURRENT MESSAGE WHEN CONNECTED TO THE CONSOLE, SIGNALING A CONDITION OCCURRED FROM WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES TO THE USER ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
234498 CORE UNIVERSAL DRIVER DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE ERL STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1