FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3132375 · Received May 29, 2013

Report

Report Number
1416980-2013-13487
Event Type
Malfunction
Date Received
May 29, 2013
Date of Event
March 29, 2013
Report Date
May 9, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING EVALUATION OF A HOMECHOICE HARDWARE DEVICE (B)(4) AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, A SYSTEM ERROR 2240 (AIR IN SET) ALARM WAS IDENTIFIED IN THE LOG WHICH INDICATES A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE. THE ALARM OCCURRED ON (B)(6) 2013 AT 01:25:09. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
234896 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 52 YR HOMECHOICE