FDA Adverse Event Malfunction Summary report: N

QUANTUM? MAVERICK?

MDR report key: 3132314 · Received May 29, 2013

Report

Report Number
2134265-2013-03688
Event Type
Malfunction
Date Received
May 29, 2013
Date of Event
May 1, 2013
Report Date
May 2, 2013
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR.: THE QUANTUM MAVERICK CATHETER WAS RECEIVED INSIDE A CARRIER TUBE (HOOP). THERE WAS BLOOD IN THE BALLOON AND INFLATION LUMEN. MAGNIFIED INSPECTION REVEALED A PINHOLE IN THE BALLOON WALL 3MM FROM THE DISTAL TIP. THE BALLOON PRESENTED NO IRREGULARITIES IN THE BALLOON MATERIAL OR THE RO MARKER THAT COULD HAVE CONTRIBUTED TO THE DAMAGE. THERE WAS NO EVIDENCE OF ANY PRODUCT QUALITY DEFICIENCIES CONTRIBUTING TO THE DAMAGE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION, BALLOON RUPTURE OCCURRED. THE STENOSED TARGET LESION WAS CALCIFIED. A 3.0MM X 8MM QUANTUM MAVERICK BALLOON CATHETER WAS USED TO PREDILATE THE TARGET LESION. UPON FIRST INFLATION THE BALLOON RUPTURED BELOW THE RATED BURST PRESSURE. THE BALLOON WAS REMOVED INTACT FROM THE PATIENT'S BODY. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE NOTED AND THE PATIENT'S CONDITION WAS GOOD.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION, BALLOON RUPTURE OCCURRED. THE STENOSED TARGET LESION WAS CALCIFIED. A 3.0MM X 8MM QUANTUM MAVERICK BALLOON CATHETER WAS USED TO PREDILATE THE TARGET LESION. UPON FIRST INFLATION THE BALLOON RUPTURED BELOW THE RATED BURST PRESSURE. THE BALLOON WAS REMOVED INTACT FROM THE PATIENT'S BODY. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE NOTED AND THE PATIENT'S CONDITION WAS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
236413 QUANTUM? MAVERICK? CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493808008300 15486537

Patients

Seq Age Sex Outcome Treatment
1