FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3132165 · Received May 29, 2013

Report

Report Number
1416980-2013-13539
Event Type
Malfunction
Date Received
May 29, 2013
Date of Event
March 19, 2013
Report Date
May 3, 2013
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4).VISUAL INSPECTION AND FUNCTIONAL TESTS WERE PERFORMED AND THERE WERE NO ISSUES FOUND. A SIMULATED THERAPY WAS PERFORMED ON THE DEVICE WITHOUT ISSUES. THE REPORTED ISSUE COULD NOT BE CONFIRMED OR DUPLICATED DURING EVALUATION OR IN THE DEVICE LOG. A REVIEW OF THE SERVICE HISTORY REVEALED NO FAILURES OR PROBLEMS THAT WERE THE SAME AS, OR SIMILAR TO, THE CURRENT DIFFICULTY AND THERE WAS NO INDICATION THAT THE PARTS REPLACED DURING SERVICING CAUSED OR CONTRIBUTED TO THE REPORTED DIFFICULTY. A REVIEW OF THE DEVICE HISTORY REVEALED NO ISSUES THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED DIFFICULTY. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. UPON COMPLETION OF THE DEVICE EVALUATION, OR IF AN ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT PERFORMING THERAPY USING A HOMECHOICE (HC) DEVICE RECEIVED A FILL VOLUME LARGER THAN THE PRESCRIBED VOLUME (200ML) DURING THE LAST FILL. THE ADMINISTERED VOLUME EQUALED 201ML. THERE WAS PATIENT INVOLVEMENT, BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
235686 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1