ETRIO
Report
- Report Number
- 3005075853-2013-02602
- Event Type
- Malfunction
- Date Received
- May 29, 2013
- Date of Event
- May 8, 2013
- Report Date
- May 10, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GEI
- PMA / PMN Number
- K072177
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RECEIVED WITH THE ELECTRODE SEPARATED FROM THE CERAMIC AND THE ELECTRODE IS STILL ATTACHED TO THE ACTIVE ROD. TESTING FOUND A SHORT ON THE DEVICE WHEN THE JAWS ARE FULLY OR PARTIALLY CLOSED, RESULTING IN A YELLOW MESSAGE SCREEN "REPOSITION JAWS AND REACTIVE" IS ADVISING THAT THE INSTRUMENT IS BEING ACTIVATED ON LOW IMPEDANCE (THIN) TISSUE OR METAL (SUCH AS STAPLES, CLIPS, RETRACTORS, OR CLAMPS). THE "REPLACE INSTRUMENT" SCREEN WOULD BE DISPLAYED AFTER RECEIVING THE "REPOSITION AND REACTIVATE" MESSAGE TWICE.
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC ESOPHAGECTOMY, THE ELECTRODE PEELED AWAY AT THE END OF OPERATION. NO PIECES FELL INTO THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ONE DEVICE WILL BE RETURNING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 236790 | ETRIO | ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES | GEI | ETHICON ENDO-SURGERY, LLC. | UNK | K4C6OA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR AND HANDPIECE |