OMNIPOD INSULIN PUMP
Report
- Report Number
- 3004464228-2013-00399
- Event Type
- Malfunction
- Date Received
- April 30, 2013
- Date of Event
- April 1, 2013
- Report Date
- April 1, 2013
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042792
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS NOT RETURNED FOR EVALUATION. WE ARE UNABLE TO CONFIRM THE KINKED CANNULA OR TO DETERMINE IF IT COULD HAVE CONTRIBUTED TO THE REPORTED HYPERGLYCEMIA. LOT QUALIFICATION RECORDS WERE REVIEWED, AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. THE OMNIPOD USER GUIDE WARNS, "TEST RESULTS GREATER THAN 250 MG/DL MEAN HIGH BLOOD GLUCOSE (HYPERGLYCEMIA). IF YOU GET RESULTS ABOVE 250 MG/DL, BUT DO NOT HAVE SYMPTOMS OF HYPERGLYCEMIA, REPEAT THE TEST. IF YOU HAVE SYMPTOMS OR CONTINUE TO GET RESULTS THAT FALL ABOVE 250 MG/DL, FOLLOW THE TREATMENT ADVICE OF YOUR HEALTHCARE PROVIDER." IT ADVISES, "IF YOUR BLOOD GLUCOSE IS 250 MG/DL OR ABOVE, CHECK FOR KETONES. IF KETONES ARE PRESENT, FOLLOW YOUR HEALTHCARE PROVIDER'S GUIDES. IF KETONES ARE NOT PRESENT, TAKE A CORRECTION BOLUS AS PRESCRIBED BY YOUR HEALTHCARE PROVIDER. CHECK BLOOD GLUCOSE AGAIN AFTER 2 HOURS. IF BLOOD GLUCOSE LEVELS HAVE NOT DECREASED, TAKE A SECOND BOLUS BY INJECTION, USING A STERILE SYRINGE. IF BLOOD GLUCOSE REMAINS HIGH AFTER ANOTHER 2 HOURS (A TOTAL OF 4 HOURS), REPLACE THE POD" AND "TO AVOID HYPERGLYCEMIA (HIGH BLOOD GLUCOSE), CHECK YOUR BLOOD GLUCOSE AT LEAST 4-6 TIMES A DAY (WHEN YOU WAKE UP, BEFORE EACH MEAL, AND BEFORE GOING TO BED."
THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE WAS 495 MG/DL. WHEN SHE REMOVED THE POD, SHE NOTICED THAT THE CANNULA WAS KINKED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 188967 | OMNIPOD INSULIN PUMP | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 11200 | L31005 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR |