2520274-2013-02920
Report
- Report Number
- 2520274-2013-02920
- Event Type
- Malfunction
- Date Received
- May 29, 2013
- Date of Event
- April 25, 2013
- Report Date
- April 30, 2013
- Manufacturer
- SYNTHES USA
- Product Code
- HWC
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER
Narratives
THIS DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. ADD'L PRO CODE: HRS. ADDITIONAL INFORMATION: IT IS UNKNOWN WHICH DEVICE THE CHIP CAME FROM. THE DEVICE WAS RECEIVED, THE INVESTIGATION COULD NOT BE COMPLETED, AND NO CONCLUSION COULD BE DRAWN, AS PRODUCT IS ENTERING THE COMPLAINT SYSTEM. AS NO PART OR LOT NUMBER WAS GIVEN FOR THE DEVICE RETURNED, A DEVICE HISTORY RECORDS REVIEW CAN NOT BE COMPLETED. THE INVESTIGATION IS ONGOING.
ADDITIONAL NARRATIVE: THIS DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. VISUAL INSPECTION OF THE METAL PART INDICATES THAT THIS MUST BE SHAVING FROM A LOCKING SCREW WHICH SHARED OFF DURING INSERTION. NO FURTHER INVESTIGATION IS POSSIBLE. WE ARE NOT ABLE TO IDENTIFY ANY ROOT CAUSE FOR THE REPORTED PROBLEM. MOST PROBABLY REASON MAY BE STRONG CONTACT WITH THE PLATE DURING INSERTION PROCEDURE. PLEASE NOTE, THE 2.7MM VA LOCKING SCREW IS NOT COMPATIBLE WITH TWO-COLUMN PLATE. PLACEHOLDER.
A DEVICE REPORT FROM SYNTHES (B)(4) INDICATED A SURGEON IN (B)(4) REPORTED: DURING SURGERY FOR A DISTAL RADIUS FRACTURE, THE SURGEON INSERTED AND LOCKED 4 VA LOCKING SCREWS IN DISTAL HOLES BY GUIDING BLOCK, AND FIXED BY DRIVER WITH TORQUE LIMITER. SECOND, HE INSERTED AND LOCKED 2.7MM 2 CORT.SCREWS IN PROXIMAL SHAFT HOLES BY DRIVER WITH TORQUE LIMITER. THIRD, HE INSERTED AND LOCKED 2.7MM 2 VA LOCKING SCREWS IN PROXIMAL VA HOLES BY TORQUE LIMITER, BY THIS TIME, ALL SCREWS COULD BE LOCKED. AND THEN HE TOOK OFF GUIDING BLOCK FIXED VA LOCKING SCREWS BY DRIVE WITH TORQUE LIMITER JUST IN CASE. HOWEVER, VA LOCKING SCREW IN DISTAL ROW MOST STYLOID SIDE HAD NOT STOPPED ROTATION. FINALLY THIS SCREW WAS PENETRATED IN HOLE. HE FOUND A METAL PART LIKE THREAD WHICH WAS ADHERED IN THE PATIENT BODY. SURGERY WAS PERFORMED BY GLOBAL TECHNICAL METHOD. HE CHOSE FIXED MODE, AND USED TORQUE LIMITER 0.8NM IN LOCK. THE METAL CHIP IS BEING RETURNED. THIS IS 1 OF 3 REPORTS FOR COMPLAINT (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 236636 | HWC | SYNTHES USA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR |