FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3131879 · Received May 29, 2013

Report

Report Number
2531779-2013-07161
Event Type
Malfunction
Date Received
May 29, 2013
Report Date
May 3, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS BUT EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION FOLLOW-UP #1 SUBMITTED 07/24/2013: THE DEVICE WAS RETURNED TO ANIMAS AND EVALUATED BY PRODUCT ANALYSIS ON 06/26/2013 WITH THE FOLLOWING RESULTS: TESTING CONFIRMED THE DISPLAY WAS FADED AND PINK. REMOVED DISPLAYED AND PLACED NEW GOOD DISPLAY IN. NEW DISPLAY CONTRAST RETURNED TO NORMAL. UNRELATED TO THIS COMPLAINT, THE BATTERY COMPARTMENT WAS CRACKED FROM THE THREADS TO CASE SEAL. BATTERY COMPARTMENT AND CAP THREADS WERE INTACT. ABLE TO FULLY TIGHTEN BATTERY CAP. OPENED PUMP. NO CORROSION OR MOISTURE IN PUMP.

Description of Event or Problem · 1

ON (B)(6) 2013, THE PATIENT CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
235400 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 42 YR