EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM
Report
- Report Number
- 2024168-2013-03320
- Event Type
- Malfunction
- Date Received
- May 29, 2013
- Date of Event
- May 3, 2013
- Report Date
- May 3, 2013
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- NTE
- PMA / PMN Number
- K090665
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE FOREIGN MATERIAL WAS UNABLE TO BE CONFIRMED. BASED ON A VISUAL INSPECTION AND SCANNING ELECTRON MICROSCOPY IMAGING OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD AND COMPLAINT HISTORY OF THE REPORTED LOT COULD NOT BE CONDUCTED BECAUSE THE LOT NUMBER WAS NOT PROVIDED. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.
(B)(4). THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.
IT WAS REPORTED, WHEN THE EMBOSHIELD NAV6 WAS REMOVED FROM THE ANATOMY AFTER USE, THERE WAS A SUBSTANCE ON THE FILTER. IT APPEARED TO BE LINT OR SHEARED MESH FROM THE FILTER. THERE WERE NO ADVERSE PATIENT EFFECT AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.