FDA Adverse Event Malfunction Summary report: N

CORDLESS DRIVER 3

MDR report key: 3131830 · Received May 29, 2013

Report

Report Number
0001811755-2013-01228
Event Type
Malfunction
Date Received
May 29, 2013
Date of Event
May 1, 2013
Report Date
May 2, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
HRX
PMA / PMN Number
K943323
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING THE DEVICE EVALUATION, THE REAR MOTOR WINDING WAS FOUND TO HAVE A SHORT CIRCUIT. A SHORT CIRCUIT CAN CAUSE AN INCREASED CURRENT DRAW, WHICH CAN RESULT IN OVERHEATING.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE AT THE USER FACILITY, THE HANDPIECE WAS OVERHEATING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
235699 CORDLESS DRIVER 3 ARTHROSCOPE HRX STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1