FDA Adverse Event Malfunction Summary report: N

STEALTHSTATION S7 SYSTEM

MDR report key: 3131789 · Received May 29, 2013

Report

Report Number
1723170-2013-00383
Event Type
Malfunction
Date Received
May 29, 2013
Date of Event
May 6, 2013
Report Date
May 6, 2013
Manufacturer
MEDTRONIC NAVIGATION, INC.
Product Code
HAW
PMA / PMN Number
K050438
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

PATIENT WEIGHT UNAVAILABLE FROM SITE AT TIME OF THIS REPORT. A MEDTRONIC REPRESENTATIVE AT THE SITE, FOUND THE VERTEK ARM WAS ABLE TO BE TIGHTENED AND HOLD IN PLACE PROPERLY; FULLY FUNCTIONAL. SOFTWARE INVESTIGATION COMPLETED. FINDINGS ARE THAT SOFTWARE ENGINEERS WERE UNABLE TO DETERMINE ROOT CAUSE AS THE REPORTED BEHAVIOR COULD NOT BE REPLICATED. THE SURGEON NOTED THE FRAME COULD HAVE BEEN PLACED IMPROPERLY OR THE VERTEK ARM MAY HAVE MOVED DURING DRAPING. THE RE-SCHEDULED CASE WAS SUCCESSFUL.

Description of Event or Problem · 1

A MEDTRONIC REPRESENTATIVE REPORTED THAT A SURGEON ALLEGED AN INACCURACY, WHILE IN A TUMOR RESECTION, USING SYNERGY CRANIAL. TRACER REGISTRATION PASSED AND ACCURACY WAS VERIFIED. AFTER REMOVING THE UNSTERILE FRAME, DRAPING AND PUTTING ON THE STERILE FRAME, THE BONE FLAP WAS REMOVED BEFORE CHECKING ACCURACY AGAIN. AFTER TAKING OFF THE BONE FLAP, THE PASSIVE PLANAR BLUNT WAS USED, THE SURGEON TOUCHED THE BONE AND THE POINTER SHOWED DEEP INSIDE THE SKULL (A MILLIMETER MEASUREMENT COULD NOT BE GIVEN). IT WAS NOTED THAT SOMETIMES WHEN TRYING TO TRACK THE PROBE, THE 2D IMAGES DID NOT APPEAR IN THEIR QUADRANTS. THE SURGEON OPTED TO DISCONTINUE THE USE OF NAVIGATION AND RE-SCHEDULED THE PROCEDURE FOR THE FOLLOWING DAY. THERE WAS NO HARM TO THE PATIENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
234723 STEALTHSTATION S7 SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT HAW MEDTRONIC NAVIGATION, INC. S7

Patients

Seq Age Sex Outcome Treatment
1 48 YR Death