ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-07157
- Event Type
- Malfunction
- Date Received
- May 29, 2013
- Report Date
- May 1, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP #1 DATE OF SUBMISSION 07/23/2013-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION REVEALED THAT THE UP ARROW AND DOWN ARROW KEYPAD BUTTONS ARE INTERMITTENTLY RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD CONTAMINATION FOUND UNDER ALL KEY CONTACTS.
THE PUMP HAS BEEN RETURNED TO ANIMAS BUT EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.
ON (B)(6) 2013, IT WAS REPORTED THAT THE KEYPAD BUTTONS WERE STICKING. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE REMAINED UNRESOLVED AT THE TIME OF TROUBLE SHOOTING. THERE IS NO INDICATION THAT THE PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 236687 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR |