SITTER SELECT
Report
- Report Number
- 2020362-2013-00210
- Event Type
- Malfunction
- Date Received
- May 6, 2013
- Report Date
- April 22, 2013
- Manufacturer
- J. T. POSEY CO.
- Product Code
- KMI
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
RESULTS: EVALUATION OF THE RETURNED PRODUCT DID NOT CONFIRM THE REPORTED ISSUE; BOTH THE VOICE AND TONE SOUND AS IT SHOULD, NO STATIC IS HEARD. HOWEVER, THE BATTERY SPRINGS INSIDE THE BATTERY COMPARTMENT ARE BADLY BENT. NO INTERMITTENCY OBSERVED DURING EVALUATION. THE BATTERY DOOR IS CHIPPED BUT CLOSES SECURELY. UNIT PASSES ALL FUNCTIONAL TESTS. NOTE: INSTRUCTIONS FOR USE HAS A STATEMENT ABOUT BATTERY REPLACEMENT: SLIDE BATTERY DOOR OPEN AND FLIP IT UP. DO NOT ATTEMPT TO REMOVE BATTERY DOOR; IT DOES NOT SEPARATE FROM ALARM. CAREFULLY REMOVE BATTERIES TO AVOID DAMAGE TO BATTERY DOOR. HOLD ALARM VERTICALLY UPSIDE-DOWN TO PREVENT BATTERY SPRING FROM BENDING DURING BATTERY INSTALLATION. INSERT FOUR (4) NEW "AA ALKALINE BATTERIES. TAKE CARE NOT TO DAMAGE BATTERY CONTACTS. FLIP BATTERY DOOR DOWN. PUSH DOWN GENTLY ON BATTERIES AND SLID BATTERY DOOR CLOSED UNTIL IT CLICKS SHUT. (B)(4).
CUSTOMER REPORTED THE ALARM HAS POWER BUT DOES NOT SOUND WHEN THE OPTION TONE AND VOICE IS USED. THE CUSTOMER ONLY HEARS STATIC. CUSTOMER ALSO REPORTED THE ALARM DOES NOT WORK WITH THE TONE OPTION EITHER. THE BATTERIES HAVE BEEN REPLACED AND THERE IS NO VISIBLE DAMAGE TO THE OUTSIDE OF THE ALARM. THE CUSTOMER DID NOT PROVIDE THE DATE WHEN THIS WAS FOUND BUT REPORTED IT WAS DISCOVERED DURING SET UP. NO PATIENT INCIDENT OR INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 197909 | SITTER SELECT | KMI | J. T. POSEY CO. | 8361 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | SITTER SELECT MAGNET: MODEL 8361, LOT # NA |