FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 880I SYNCHRON ® ACCESS CLINICAL SYSTEM

MDR report key: 3131652 · Received May 29, 2013

Report

Report Number
2050012-2013-00364
Event Type
Malfunction
Date Received
May 29, 2013
Date of Event
April 26, 2013
Report Date
April 29, 2013
Manufacturer
BECKMAN COULTER
Product Code
JJE
PMA / PMN Number
K060256
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

A BECKMAN COULTER FSE (FIELD SERVICE ENGINEER) WAS DISPATCHED AND REPLACED THE CREM MODULE, AND CALIBRATED THE LAMP AND CUP TO RESOLVE THE ISSUE. REPAIR WAS VERIFIED PER ESTABLISHED PROCEDURES AND RESULTS MET PUBLISHED SPECIFICATIONS. FAILURE MODE APPEARS TO BE THE CREATININE MODULE AND REPLACEMENT OF THE MODULE RESOLVED THE ISSUE. RESULTS: CREATININE MODULE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED ON (B)(4) 2013 THAT THE STIR BAR IN THE CREATININE CUP MODULE OF THE UNICEL DXC 880I SYNCHRON ACCESS CLINICAL SYSTEM WAS NOT TURNING. UPON FOLLOW-UP BY A BECKMAN COULTER CTS (CUSTOMER TECHNICAL SUPPORT) ON (B)(4) 2013, THE CUSTOMER STATED THAT TWO (2) FALSE HIGH CREM (MODULAR CREATININE) RESULTS WERE GENERATED AND REPORTED OUTSIDE OF THE LABORATORY. THE CUSTOMER REPEATED THE PATIENT SAMPLES ON AN ALTERNATE DXC INSTRUMENT AND ISSUED AMENDED REPORTS. THERE WAS NO CHANGE OR AFFECT TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT. LEVEL 3 QC (QUALITY CONTROL) RESULTS EXHIBITED IMPRECISION. NO OTHER CHEMISTRIES WERE AFFECTED DURING THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
237318 UNICEL® DXC 880I SYNCHRON ® ACCESS CLINICAL SYSTEM ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1