FDA Adverse Event Malfunction Summary report: N

6000 CMS IOD

MDR report key: 3131594 · Received May 17, 2013

Report

Report Number
1722139-2013-00935
Event Type
Malfunction
Date Received
May 17, 2013
Date of Event
September 3, 2010
Report Date
December 11, 2010
Manufacturer
MOOG DEVICE GROUP
Product Code
FRN
PMA / PMN Number
K981816
Removal / Correction Number
Z-1868-2011
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION FOUND THAT PUMP HAD EXPERIENCED ERROR CODE 45 IN PUMP'S HISTORY. NEW PUMP SOFTWARE WAS INSTALLED. REFERENCE RECALL NUMBER Z-1868-2011.

Description of Event or Problem · 1

INFORMATION RECEIVED STATES THAT PUMP HAD EXPERIENCED ERROR CODE 45.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
219999 6000 CMS IOD FRN MOOG DEVICE GROUP 6000 CMS IOD

Patients

Seq Age Sex Outcome Treatment
1