FDA Adverse Event Malfunction Summary report: N

OMNIPOD INSULIN PUMP

MDR report key: 3131544 · Received May 16, 2013

Report

Report Number
3004464228-2013-00461
Event Type
Malfunction
Date Received
May 16, 2013
Date of Event
April 21, 2013
Report Date
April 21, 2013
Manufacturer
INSULET CORPORATION
Product Code
LZG
PMA / PMN Number
K042792
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR EVALUATION. WE ARE UNABLE TO CONFIRM THE KINKED CANNULA OR TO DETERMINE IF IT COULD HAVE CONTRIBUTED TO THE REPORTED HYPERGLYCEMIA. LOT QUALIFICATION RECORDS WERE REVIEWED, AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. THE OMNIPOD USER GUIDE WARNS, "TEST RESULTS GREATER THAN 250 MG/DL MEAN HIGH BLOOD GLUCOSE (HYPERGLYCEMIA). IF YOU GET RESULTS ABOVE 250 MG/DL, BUT DO NOT HAVE SYMPTOMS OF HYPERGLYCEMIA, REPEAT THE TEST. IF YOU HAVE SYMPTOMS OR CONTINUE TO GET RESULTS THAT FALL ABOVE 250 MG/DL, FOLLOW THE TREATMENT ADVICE OF YOUR HEALTHCARE PROVIDER." IT ADVISES, "IF YOUR BLOOD GLUCOSE IS 250 MG/DL OR ABOVE, CHECK FOR KETONES. IF KETONES ARE PRESENT, FOLLOW YOUR HEALTHCARE PROVIDER'S GUIDELINES. IF KETONES ARE NOT PRESENT, TAKE A CORRECTION BOLUS AS PRESCRIBED BY YOUR HEALTHCARE PROVIDER. CHECK BLOOD GLUCOSE AGAIN AFTER 2 HOURS. IF BLOOD GLUCOSE LEVELS HAVE NOT DECREASED, TAKE A SECOND BOLUS BY INJECTION, USING A STERILE SYRINGE. IF BLOOD GLUCOSE REMAINS HIGH AFTER ANOTHER 2 HOURS (A TOTAL OF 4 HOURS), REPLACE THE POD."

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE WAS 85 MG/DL WHEN SHE ACTIVATED A POD AT 11:30 AM. AT 12:30 PM, WITH A BLOOD GLUCOSE OF 200 MG/DL, SHE GAVE HERSELF A 2 UNIT BOLUS. AT 1:30 PM, IT WAS 340 MG/DL, AND SHE GAVE HERSELF A 2.8 BOLUS. AT 2:30 PM, SHE DEACTIVATED THE POD. SHE STATED THAT THE CANNULA LOOKED KINKED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
218179 OMNIPOD INSULIN PUMP PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 11200 L31036

Patients

Seq Age Sex Outcome Treatment
1 42 YR