FDA Adverse Event Malfunction Summary report: N

OMNIPOD INSULIN PUMP

MDR report key: 3131527 · Received May 14, 2013

Report

Report Number
3004464228-2013-00438
Event Type
Malfunction
Date Received
May 14, 2013
Date of Event
April 8, 2013
Report Date
April 15, 2013
Manufacturer
INSULET CORPORATION
Product Code
LZG
PMA / PMN Number
K042792
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS DISCARDED AND WILL NOT BE RETURNED FOR EVALUATION. WE ARE UNABLE TO DETERMINE IF ANY MALFUNCTION COULD HAVE CONTRIBUTED TO THE REPORTED HYPERGLYCEMIA. THE CUSTOMER REPORTED THAT THE CANNULA HAD DISLODGE FROM THE INFUSION SITE. THIS CONDITION COULD INTERRUPT INSULIN DELIVERY AND CONTRIBUTE TO HYPERGLYCEMIA. LOT QUALIFICATION RECORDS WERE REVIEWED, AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. THE OMNIPOD USER GUIDE WARNS "CHECK OFTEN TO MAKE SURE THE POD AND SOFT CANNULA ARE SECURELY ATTACHED AND IN PLACE. A LOOSE OR DISLODGED CANNULA MAY INTERRUPT INSULIN DELIVERY" AND "BECAUSE INSULIN PODS USE ONLY RAPID-ACTING INSULIN, USERS ARE AT INCREASED RISK FOR DEVELOPING HYPERGLYCEMIA (HIGH BLOOD GLUCOSE) IF INSULIN DELIVERY IS INTERRUPTED. IF IT IS UNTREATED, SEVERE HYPERGLYCEMIA CAN QUICKLY LEAD TO DIABETIC KETOACIDOSIS (DKA). DKA CAN CAUSE BREATHING DIFFICULTIES, SHOCK, COMA, OR DEATH. IF INSULIN DELIVERY IS INTERRUPTED FOR ANY REASON, YOU MAY NEED TO REPLACE THE MISSING INSULIN - USUALLY WITH AN INJECTION OF RAPID-ACTING INSULIN. ASK YOUR HEALTHCARE PROVIDER FOR INSTRUCTIONS ON HANDLING INTERRUPTED INSULIN DELIVERY." IT ALSO WARNS, "TEST RESULTS GREATER THAN 250 MG/DL MEAN HIGH BLOOD GLUCOSE (HYPERGLYCEMIA). IF YOU GET RESULTS ABOVE 250 MG/DL, BUT DO NOT HAVE SYMPTOMS OF HYPERGLYCEMIA, REPEAT THE TEST. IF YOU HAVE SYMPTOMS OR CONTINUE TO GET RESULTS THAT FALL ABOVE 250 MG/DL, FOLLOW THE TREATMENT ADVICE OF YOUR HEALTHCARE PROVIDER."

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT HIS BLOOD GLUCOSE RESULT WAS 412 MG/DL AT 8:46 AM, WHICH HE CORRECTED WITH A 5.5 UNIT BOLUS. AT 9:24 AM, HIS RESULT WAS 421 MG/DL, AND HE GAVE HIMSELF A 2 UNIT BOLUS. HE REMOVED THE POD AND NOTICED THAT THE CANNULA HAD DISLODGED FROM THE INFUSION SITE. HE DISPOSED OF THE POD PRIOR TO CALLING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
213867 OMNIPOD INSULIN PUMP PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 11200 L31069

Patients

Seq Age Sex Outcome Treatment
1 15 YR