LIBERTY CYCLER SET, DUAL PATIENT CONNECT
Report
- Report Number
- 8030665-2013-00281
- Event Type
- Malfunction
- Date Received
- May 14, 2013
- Date of Event
- April 15, 2013
- Report Date
- April 15, 2013
- Manufacturer
- REYNOSA MANUFACTURING
- Product Code
- FKX
- PMA / PMN Number
- K043363
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
THE ACTUAL DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR PHYSICAL EVALUATION, HOWEVER, A COMPANION SAMPLE FROM THE SAME LOT WAS RETURNED FOR INVESTIGATION AND NO DEFECTS WERE NOTED. IN ADDITION, AN INVESTIGATION OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED BY THE MANUFACTURER. THERE WERE NO DEVIATIONS OR NONCONFORMANCE DURING THE MANUFACTURING PROCESS. IN ADDITION, THE BATCH RECORD REVIEW CONFIRMED THE LABELING, MATERIAL, AND PROCESS CONTROLS WERE WITHIN SPECIFICATION.
A PERITONEAL DIALYSIS PATIENT HAS REPORTED THAT DIALYSIS SOLUTION WAS LEAKING OUT OF THE CASSETTE AND INTO THE CYCLER. UPON REMOVING THE TUBING SET FOLLOWING TREATMENT, FLUID WAS LEAKING FROM THE CASSETTE DOOR OF THE CYCLER. PATIENT DID NOT RECEIVE AN ANTIBIOTICS AND HAD NO ADVERSE EFFECTS. A COMPANION SAMPLE OF THE SAME LOT NUMBER IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 213768 | LIBERTY CYCLER SET, DUAL PATIENT CONNECT | FKX | REYNOSA MANUFACTURING | 13AR08812 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | LIBERTY DIALYSIS CYCLER |