FDA Adverse Event Malfunction Summary report: N

LIBERTY CYCLER SET, DUAL PATIENT CONNECT

MDR report key: 3131496 · Received May 14, 2013

Report

Report Number
8030665-2013-00281
Event Type
Malfunction
Date Received
May 14, 2013
Date of Event
April 15, 2013
Report Date
April 15, 2013
Manufacturer
REYNOSA MANUFACTURING
Product Code
FKX
PMA / PMN Number
K043363
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR PHYSICAL EVALUATION, HOWEVER, A COMPANION SAMPLE FROM THE SAME LOT WAS RETURNED FOR INVESTIGATION AND NO DEFECTS WERE NOTED. IN ADDITION, AN INVESTIGATION OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED BY THE MANUFACTURER. THERE WERE NO DEVIATIONS OR NONCONFORMANCE DURING THE MANUFACTURING PROCESS. IN ADDITION, THE BATCH RECORD REVIEW CONFIRMED THE LABELING, MATERIAL, AND PROCESS CONTROLS WERE WITHIN SPECIFICATION.

Description of Event or Problem · 1

A PERITONEAL DIALYSIS PATIENT HAS REPORTED THAT DIALYSIS SOLUTION WAS LEAKING OUT OF THE CASSETTE AND INTO THE CYCLER. UPON REMOVING THE TUBING SET FOLLOWING TREATMENT, FLUID WAS LEAKING FROM THE CASSETTE DOOR OF THE CYCLER. PATIENT DID NOT RECEIVE AN ANTIBIOTICS AND HAD NO ADVERSE EFFECTS. A COMPANION SAMPLE OF THE SAME LOT NUMBER IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
213768 LIBERTY CYCLER SET, DUAL PATIENT CONNECT FKX REYNOSA MANUFACTURING 13AR08812

Patients

Seq Age Sex Outcome Treatment
1 LIBERTY DIALYSIS CYCLER