FDA Adverse Event Other Summary report: N

OLYMPUS HIGH DEFINITION LCD MONITOR

MDR report key: 3131469 · Received May 8, 2013

Report

Report Number
8010047-2013-00146
Event Type
Other
Date Received
May 8, 2013
Date of Event
February 26, 2013
Report Date
April 10, 2013
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORPORATION
Product Code
NWB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

OLYMPUS FOLLOWED-UP WITH THE USER FACILITY TO OBTAIN ADDITIONAL INFO REGARDING THIS REPORT. THE POWER SUPPLY WAS IMPROPERLY AFFIXED TO THE PLATE COVER WITH TWO ZIP TIES AND ATTACHED TO THE BOOM ARM. PER THE USER FACILITY, THE INSTALLATION WAS PERFORMED BY OLYMPUS. THE USER FACILITY INDICATED THAT THE DEVICE WAS STILL IN USE. THE USER FACILITY REPORTED THE INCIDENT TO FDA. OLYMPUS FOLLOWED-UP WITH THE FIELD SERVICE ENGINEER (FSE) TO OBTAIN ADDITIONAL INFO REGARDING THIS REPORT. THE FSE REPORTED THAT POSSIBLY LOOSE SCREWS ON THE PLATE COVER RESULTED IN THE REPORTED EVENT. NO PRODUCT WAS RETURNED TO OLYMPUS. THE MOST LIKELY CONTRIBUTING FACTOR WAS UNSATISFACTORY INSTALLATION. OLYMPUS WAS INFORMED THAT THE USER FACILITY CORRECTED THE INSTALLATION BY PLACING THE POWER SUPPLY ONTO THE CEILING AND REWIRING THE LOW VOLTAGE POWER SUPPLY FROM THE MONITOR TO PREVENT REOCCURRENCES. ALL OF THE MONITORS WITHIN THE FACILITY WERE REINSTALLED.

Description of Event or Problem · 1

OLYMPUS WAS INFORMED THAT DURING A VIDEO-ASSISTED THORACOSCOPY SURGERY (VATS) THERAPEUTIC PROCEDURE, THE PLATE COVER WITH THE ATTACHED POWER SUPPLY FELL OFF FROM THE REAR OF THE MONITOR (WHICH WAS ATTACHED TO BOOM ARM) TO THE GROUND, ALMOST HITTING A SCRUB NURSE. NO ONE WAS INJURED. THE PROCEDURE WAS SIGNIFICANTLY PROLONGED. THE PLATE COVER HAD TWO EXPOSED SCREWS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
202413 OLYMPUS HIGH DEFINITION LCD MONITOR MONITOR NWB OLYMPUS MEDICAL SYSTEMS CORPORATION OEV261H NA

Patients

Seq Age Sex Outcome Treatment
1 BOOM WITH UNK MODEL NUMBER, MFG BY TRUMP