FDA Adverse Event Malfunction Summary report: N

THERMOPHORE AUTOMATIC MOIST HEAT PACK

MDR report key: 3131461 · Received February 25, 2013

Report

Report Number
1811605-2013-00025
Event Type
Malfunction
Date Received
February 25, 2013
Date of Event
February 1, 2013
Report Date
February 19, 2013
Manufacturer
BATTLE CREEK EQUIPMENT COMPANY
Product Code
IRT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

IT WAS REPORTED THAT THE FABRIC FOUNDATION OF THE UNIT WAS BURNED BY AN INTERNAL COMPONENT. OUR EXAMINATION REVEALED THAT ONE OF THE TERMINALS HAD BEEN BROKEN NEAR A THERMOSTAT, RESULTING IN A BURN WHICH HAD COMPROMISED OUR DOUBLE-INSULATED DESIGN, CAUSING A 1/4" DIAMETER HOLE IN THE FABRIC FOUNDATION. WE ALSO FOUND SEVERAL OTHER INTERNAL COMPONENTS THAT WERE DAMAGED, INDICATING THAT THE CUSTOMER MISUSED THE PAD BY APPLYING AN EXCESSIVE, LOCALIZED FORCE ON OR NEAR THE THERMOSTAT TERMINAL. WE CONCLUDED THAT THIS REPORT WAS ATTRIBUTABLE TO MISUSE BY THE USER, AND FAILURE TO FOLLOW INSTRUCTIONS AS TO THE CARE AND HANDLING OF THE UNIT. ALTHOUGH WARNING LABELS AND INSTRUCTION BOOKLETS WARN AGAINST SITTING OR LAYING ON THE PRODUCT, WE HAVE A CORRECTIVE ACTION PROJECT (B)(4) UNDERWAY TO MAKE THE DESIGN MORE ROBUST.

Description of Event or Problem · 1

THE USER REPORTED THAT THE PAD HAD A 1/4" DIAMETER BURN HOLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
81144 THERMOPHORE AUTOMATIC MOIST HEAT PACK ELECTRIC MOIST HEAT PACK IRT BATTLE CREEK EQUIPMENT COMPANY 055

Patients

Seq Age Sex Outcome Treatment
1