FDA Adverse Event Injury Summary report: N

MEDICHOICE

MDR report key: 3131079 · Received May 23, 2013

Report

Report Number
3003753847-2013-00014
Event Type
Injury
Date Received
May 23, 2013
Report Date
February 20, 2013
Manufacturer
DALIAN GOODWOOD MEDICAL CARE LTD.
Product Code
HGD
Adverse Event
Yes
Product Problem
Yes
Report Source
Distributor report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THE MEDICAL DEVICE ASSOCIATED WITH THIS REPORT IS A NON-STERILE, RAYON-FIBER TIPPED APPLICATOR (OB/GYN) CONSISTING OF A PLASTIC STICK AND THE RAYON-FIBER BUD. IT WAS COMMUNICATED THAT THE FIBER HEAD WAS FALLING OFF AND BEING LEFT BEHIND IN THE VAGINAL WALL. THE MANUFACTURER STATED THAT IN (B)(6) 2011 THEY REDUCED THE CONCENTRATION OF GLUE WHICH WAS USED TO SECURE THE RAYON TIP TO THE PLASTIC SHAFT. THE LOW CONCENTRATION OF GLUE USED TO SECURE THE RAYON TIP TO THE PLASTIC SHAFT WAS DETERMINED TO BE THE MOST PROBABLE ROOT CAUSE OF THE FIBER HEAD FALLING OFF.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
229502 MEDICHOICE RAYON TIPPED OB/GYN APPLICATOR HGD DALIAN GOODWOOD MEDICAL CARE LTD. 1211DG13A

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention