FDA Adverse Event Injury Summary report: N

NOVASURE IMPEDANCE CONTROLLED EA SYSTEM

MDR report key: 3131021 · Received May 10, 2013

Report

Report Number
1222780-2013-00087
Event Type
Injury
Date Received
May 10, 2013
Date of Event
March 8, 2013
Report Date
April 12, 2013
Manufacturer
HOLOGIC
Product Code
MNB
PMA / PMN Number
P010013
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

SERIAL NUMBER OF THE DISPOSABLE DEVICE NOT PROVIDED BY THE COMPLAINANT. SERIAL NUMBER OF THE RADIO FREQUENCY CONTROLLER NOT PROVIDED BY THE COMPLAINANT. THE DEVICE IS NOT BEING RETURNED THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE CANNOT BE COMPLETED. DEVICE HISTORY RECORD (DHR) REVIEW WAS CONDUCTED FOR THE REPORTED LOT NUMBER. THE LOT WAS RELEASED MEETING ALL QA SPECIFICATIONS. CURRENTLY UNABLE TO ESTABLISH A RELATIONSHIP OR IMPACT TO THE REPORTED OBSERVATION. ACCORDING TO THE INSTRUCTIONS FOR USE (IFU) WARNINGS: USE CAUTION NOT TO PERFORATE THE UTERINE WALL WHEN SOUNDING, DILATING, OR INSERTING THE DISPOSABLE DEVICE. IF THE DISPOSABLE DEVICE IS DIFFICULT TO INSERT INTO THE CERVICAL CANAL, USE CLINICAL JUDGEMENT TO DETERMINE WHETHER OR NOT FURTHER DILATION IS REQUIRED. THE NOVASURE SYSTEM PERFORMS A CAVITY INTEGRITY ASSESSMENT (CIA) TEST TO EVALUATE THE INTEGRITY OF THE UTERINE CAVITY, AND SOUNDS AN ALARM WARNING OF A POSSIBLE PERFORATION PRIOR TO TREATMENT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A NOVASURE ENDOMETRIAL ABLATION THE CAVITY INTEGRITY ASSESSMENT (CIA) TEST WAS UNSUCCESSFUL. THE PHYSICIAN THEN PERFORMED A HYSTEROSCOPY AND VISUALIZED A "PERFORATION". THE PROCEDURE WAS ABORTED. A LAPAROSCOPIC SUTURE WAS PLACED AND THE PT WAS DISCHARGED HOME. A DILATATION WAS PERFORMED PRIOR TO THE ATTEMPTED ABLATION. IT IS NOT KNOWN WHEN THIS PERFORATION OCCURRED OR WHAT INSTRUMENT MAY HAVE BEEN THE CAUSE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
209422 NOVASURE IMPEDANCE CONTROLLED EA SYSTEM MNB HOLOGIC NS2000 12F23R

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention RADIO FREQUENCY CONTROLLER - SERIAL # UNK