FDA Adverse Event Other Summary report: N

CLINITEK STATUS

MDR report key: 3130987 · Received April 22, 2013

Report

Report Number
1217157-2013-00068
Event Type
Other
Date Received
April 22, 2013
Date of Event
March 18, 2013
Report Date
April 12, 2013
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC
Product Code
KQO
PMA / PMN Number
K031947
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER HAS BEEN RE-TRAINED FOR PROPER SAMPLING TECHNIQUE. CUSTOMER IS BEING PROVIDED WITH SOFTWARE THAT WILL GENERATE AN ERROR CODE (E50) ERROR INSTEAD OF MISIDENTIFYING THE STRIP.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT A MULTISTIX 10SG DIPSTICK READ ERRONEOUSLY AS A MULTISTIX PRO 10LB DIPSTICK ON THE INSTRUMENT. THERE WAS NO REPORT OF INJURY FOR THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
173303 CLINITEK STATUS CT STATUS KQO SIEMENS HEALTHCARE DIAGNOSTICS INC 301026

Patients

Seq Age Sex Outcome Treatment
1