FDA Adverse Event Malfunction Summary report: N

MONOPOLAR CURVED SCISSORS INSTRUMENT

MDR report key: 3130953 · Received May 24, 2013

Report

Report Number
2955842-2013-01847
Event Type
Malfunction
Date Received
May 24, 2013
Date of Event
September 12, 2011
Report Date
September 21, 2011
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050005
Removal / Correction Number
2955842-051613-005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS MDR IS BEING SUBMITTED FOR RETROSPECTIVE ACTIVITY PERFORMED RELATING TO FIELD ACTION NUMBER (B)(4) TO INVESTIGATE MICRO-CRACKS ON THE MONOPOLAR CURVED SCISSORS INSTRUMENT. THESE TYPES OF MICRO-CRACKS WILL NOT LEAD TO MECHANICAL FAILURE OF THE INSTRUMENT; HOWEVER, THERE IS A POTENTIAL FOR INSULATION FAILURE AFTER REPROCESSING, RESULTING IN A PATHWAY FOR ELECTROSURGICAL ENERGY TO LEAK TO TISSUE AND POTENTIALLY CAUSE UNINTENDED INJURIES. THE LOCATION OF THESES MICRO-CRACKS IS CONFINED TO (B)(4) OF THE DISTAL END OF THE INSTRUMENT SHAFT. THIS INSTRUMENT WAS INSPECTED AS PART OF THIS RETROSPECTIVE ACTIVITY AND FOUND TO CONTAIN CRACKS ON THE INSTRUMENT MAIN TUBE. ENGINEERING EVALUATION WAS UNABLE TO FIND ANY EVIDENCE OF ARCING OR MELTED MATERIALS. THE INSTRUMENT WAS PLACED ON AN IN HOUSE TEST SYSTEM FOR PERFORMANCE TESTING AND NO TROUBLE WAS FOUND. THE TIP COVER ACCESSORY USED IN CONJUNCTION WITH THE MCS INSTRUMENT DURING THE PROCEDURE WAS DISCARDED BY THE HOSPITAL AND WILL NOT BE RETURNED. ADDITIONALLY, THIS EVENT WAS INITIALLY REPORTED VIA MEDWATCH REPORT NUMBER 2955842-2011-00330 ON (B)(4) 2011.

Description of Event or Problem · 1

IT WAS REPORTED THAT POST A SUCCESSFUL DA VINCI SI PROSTATECTOMY PROCEDURE, THE PATIENT RETURNED TO THE HOSPITAL EXPERIENCING COMPLICATIONS. IT WAS DISCOVERED THAT THE PATIENT HAD A PERFORATION IN HIS RIGHT COLON, APPROXIMATELY 6 CM FROM THE CAECUM. THE MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT WAS INSPECTED PRIOR TO USE. ADDITIONALLY, NO COMPLICATIONS WERE OBSERVED BY THE SURGEON OR OPERATING ROOM STAFF WITH THE MCS INSTRUMENT OR THE MCS TIP COVER ACCESSORY DURING THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
232208 MONOPOLAR CURVED SCISSORS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420179-10 M10110425 234

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SURG SYST., MCS TIP CVR, ACCSS & ESU.