FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3130860 · Received May 24, 2013

Report

Report Number
3004209178-2013-94073
Event Type
Injury
Date Received
May 24, 2013
Date of Event
May 7, 2013
Report Date
May 9, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING INCLUDING THE DISPLACEMENT TEST. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED BEING HOSPITALIZED FOR THREE DAYS DUE TO HIGH BLOOD GLUCOSE OF 800MG/DL. THE CALLER STATED THAT THE DOCTOR BELIEVES THE INSULIN PUMP WAS MALFUNCTIONING. TROUBLESHOOTING WAS PERFORMED. THE TIME, DATE, BASAL RATES, AND BOLUS WIZARD SETTINGS WERE CORRECT. THE DRIVE SUPPORT CAP APPEARS TO BE NORMAL. ASSISTED THE CUSTOMER TO RUN A MANUAL PRIME AND THE INSULIN DID EXIT. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
231103 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 89 YR Hospitalization