FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3130806 · Received May 24, 2013

Report

Report Number
3004209178-2013-94091
Event Type
Malfunction
Date Received
May 24, 2013
Date of Event
May 10, 2013
Report Date
May 10, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NM
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP IS NOT DELIVERING INSULIN. CUSTOMER STATED THAT THE BLOOD GLUCOSE LEVELS ARE HIGH. CUSTOMER'S BLOOD GLUCOSE IS 361 MG/DL. CUSTOMER HAS TREATED WITH A BOLUS AND MANUAL INJECTIONS. DURING TROUBLESHOOTING, HIGH PRESSURE TEST COMPLETED, THE INSULIN PUMP DID NOT ALARM NO DELIVERY, NO INSULIN EXITED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
232380 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-522NAH

Patients

Seq Age Sex Outcome Treatment
1 54 YR