FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3130805 · Received May 24, 2013

Report

Report Number
3004209178-2013-94088
Event Type
Injury
Date Received
May 24, 2013
Date of Event
May 1, 2013
Report Date
May 10, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

CUSTOMER'S MOTHER, (B)(6) CALLED TO REPORT A HOSPITALIZATION DUE TO HIGH BLOOD GLUCOSE. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE EVENT WAS OVE 400 MG/DL. CUSTOMER WAS VOMITING. CUSTOMER WAS DIAGNOSED WITH DIABETIC KETOACIDOSIS. TREATED WITH INSULIN DRIP. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
231667 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAP

Patients

Seq Age Sex Outcome Treatment
1 15 YR Hospitalization