FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3130796 · Received May 24, 2013

Report

Report Number
3004209178-2013-94086
Event Type
Malfunction
Date Received
May 24, 2013
Date of Event
May 10, 2013
Report Date
May 10, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP RECEIVED WITH PRIME ALARMS DURING BASIC OCCLUSION TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP HAS ALARMED PRIME ALARM. CUSTOMER'S BLOOD GLUCOSE READING IS 356 MG/DL. INSULIN SQUIRTS OUT DURING THE MANUAL PRIME PROCESS. THE DRIVE SUPPORT CAP IS PROTRUDED. ADVISED CUSTOMER THAT THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
232231 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-522NAL

Patients

Seq Age Sex Outcome Treatment
1 50 YR