FDA Adverse Event Malfunction Summary report: N

RESERVOIR 1.8ML

MDR report key: 3130791 · Received May 24, 2013

Report

Report Number
3004209178-2013-94078
Event Type
Malfunction
Date Received
May 24, 2013
Date of Event
May 10, 2013
Report Date
May 10, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
FRN
PMA / PMN Number
K001828
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSPECTED ONE OPENED AND USED RESERVOIR. PERFORMED MANUAL PRIME AND HIGH PRESSURE TEST PER SPECIFICATIONS. RAN PRIMING IN INSULIN PUMP. RESERVOIR PASSED PER SPECIFICATION. NO LEAKAGE ANOMALY WAS OBSERVED DURING ANALYSIS. CHECKED O-RINGS FOR DEFECTS AND NONE WERE FOUND.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RESERVOIR LEAKED. CUSTOMER HAS BEEN EXPERIENCING HIGH BLOOD GLUCOSE. CUSTOMER'S BLOOD GLUCOSE READING IS 343 MG/DL. CUSTOMER HAS TREATED WITH MANUAL INJECTION. THE LEAK IS PAST BOTH O-RINGS. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
232122 RESERVOIR 1.8ML RESERVOIR FRN MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-326A

Patients

Seq Age Sex Outcome Treatment
1 39 YR