FDA Adverse Event Malfunction Summary report: N

PFNA BLADE PERF L95 TAN

MDR report key: 3130662 · Received May 24, 2013

Report

Report Number
8030965-2013-02349
Event Type
Malfunction
Date Received
May 24, 2013
Date of Event
April 26, 2013
Report Date
April 27, 2013
Manufacturer
SYNTHES GMBH
Product Code
HSB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. DEVICE IS NOT DISTRIBUTED IN THE UNITED STATES, BUT IS SIMILAR TO A DEVICE MARKETED IN THE USA. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. INVESTIGATION IS ON GOING; NO CONCLUSION COULD BE DRAWN.

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS SENT TO OUR RELEVANT PM FOR INVESTIGATION: THE BLADE IS COMPLETELY JAMMED UP AND COULD ONLY BE LOOSENED UP BY USING EXCESSIVE FORCE. THIS GIVES US A CLEAR INDICATION THAT THE PFNA BLADE HAS NOT BEEN INSERTED CORRECTLY, WRONG ROTATIONAL DIRECTION. THE END CAP OF THE BLADE WAS FOUND BROKEN. BASED ON THESE FINDINGS WE CONCLUDE THAT THE CAUSE OF FAILURE IS NOT DUE TO ANY MANUFACTURING NON-CONFORMANCES. THE RETURNED ARTICLE WAS ANALYZED FOR CONFORMANCE TO PRINT SPECIFICATIONS, AS WELL AS THE DEVICE HISTORY RECORD HAS BEEN RESEARCHED. NO ABNORMAL FINDINGS WERE IDENTIFIED.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES (B)(6) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: DURING A PROCEDURE ON (B)(6) 2013, THE PROXIMAL FEMORAL NAIL A (PFNA) BLADE STUCK TO THE BLADE INSERTER. THE SURGEON FOUND IT DIFFICULT TO LOCK THE BLADE AND RELEASE THE INSERTER. THE SURGEON DECIDED TO TAKE OUT THE BLADE AND PUT IN A NEW ONE. IT WAS REPORTED THAT BECAUSE THE SURGEON WAS UNABLE TO RELEASE THE BLADE FROM THE INSERTER, HE DECIDED TO BREAK THE BLADE TO GET THE INSERTER FREE. THE TIP OF THE BLADE REMAINED ATTACHED TO THE INSERTER. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
232482 PFNA BLADE PERF L95 TAN HSB SYNTHES GMBH 8253125

Patients

Seq Age Sex Outcome Treatment
1 76 YR