FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3130609 · Received May 24, 2013

Report

Report Number
1416980-2013-13424
Event Type
Malfunction
Date Received
May 24, 2013
Date of Event
May 9, 2013
Report Date
May 9, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS NOT AVAILABLE FOR EVALUATION. IF ADDITIONAL OR RELEVANT INFORMATION BECOMES AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A HOME PATIENT (HP) HAD RECEIVED A SYSTEM ERROR (SE) 2240 ALARM (AIR IN LINE) ON THE HOME CHOICE (HC) MACHINE DURING DWELL 4 OF 6. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED WITH TROUBLESHOOTING AND CLEARED THE ALARM. THE TSR ALSO ADVISED THE HP TO START OVER WITH ALL NEW SUPPLIES OR CONTINUE WITH MANUAL BAGS. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
232193 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 73 YR HOMECHOICE