ABSORB BIORESORBABLE VASCULAR SCAFFOLD SYSTEM
Report
- Report Number
- 2024168-2013-03270
- Event Type
- Malfunction
- Date Received
- May 24, 2013
- Date of Event
- May 2, 2013
- Report Date
- May 2, 2013
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- NIQ
- PMA / PMN Number
- P070015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HK
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION. THOUGH THE DEVICE IS NOT APPROVED FOR SALE IN THE US, IT USES A DELIVERY SYSTEM WHICH IS SIMILAR TO A DEVICE SOLD IN THE US.
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE REPORTED BALLOON MATERIAL RUPTURE WAS CONFIRMED. BASED ON THE VISUAL AND FUNCTIONAL ANALYSIS OF THE DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. A REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.
IT WAS REPORTED THAT VIA A RADIAL APPROACH, THE 3.5X28 MM DEVICE WAS INFLATED SUCCESSFULLY TO 9 ATMOSPHERES (ATM) 2 TIMES AND THE IMPLANT WAS DEPLOYED SUCCESSFULLY. POST-DILATATION WAS PERFORMED WITH A NC TREK 3.5X12 MM BALLOON CATHETER. THE PROCEDURE WAS COMPLETED. THE PHYSICIAN FOUND BLOOD FILL AT THE CONNECTION POINT TO INDEFLATOR. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 231822 | ABSORB BIORESORBABLE VASCULAR SCAFFOLD SYSTEM | DRUG ELUTING CORONARY STENT SYSTEM | NIQ | AV-TEMECULA-CT | 3010761 |
Patients