FDA Adverse Event Malfunction Summary report: N

ABSORB BIORESORBABLE VASCULAR SCAFFOLD SYSTEM

MDR report key: 3130432 · Received May 24, 2013

Report

Report Number
2024168-2013-03270
Event Type
Malfunction
Date Received
May 24, 2013
Date of Event
May 2, 2013
Report Date
May 2, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P070015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HK
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION. THOUGH THE DEVICE IS NOT APPROVED FOR SALE IN THE US, IT USES A DELIVERY SYSTEM WHICH IS SIMILAR TO A DEVICE SOLD IN THE US.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE REPORTED BALLOON MATERIAL RUPTURE WAS CONFIRMED. BASED ON THE VISUAL AND FUNCTIONAL ANALYSIS OF THE DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. A REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT VIA A RADIAL APPROACH, THE 3.5X28 MM DEVICE WAS INFLATED SUCCESSFULLY TO 9 ATMOSPHERES (ATM) 2 TIMES AND THE IMPLANT WAS DEPLOYED SUCCESSFULLY. POST-DILATATION WAS PERFORMED WITH A NC TREK 3.5X12 MM BALLOON CATHETER. THE PROCEDURE WAS COMPLETED. THE PHYSICIAN FOUND BLOOD FILL AT THE CONNECTION POINT TO INDEFLATOR. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
231822 ABSORB BIORESORBABLE VASCULAR SCAFFOLD SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 3010761

Patients

Seq Age Sex Outcome Treatment
1