CUSTOM PAK
Report
- Report Number
- 3002037047-2013-00039
- Event Type
- Malfunction
- Date Received
- May 22, 2013
- Date of Event
- February 21, 2013
- Report Date
- April 25, 2013
- Manufacturer
- ALCON - BELGIUM/S.A. ALCON-COUVREUR N.V.
- Product Code
- KYG
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- UNKNOWN
Narratives
NO SAMPLES WERE RETURNED FOR EVAL; THEREFORE, THE CONDITION OF THE PRODUCT COULD NOT BE VERIFIED. THE DEVICE HISTORY RECORD (DHR) FOR THE LOT WAS REVIEWED. FUNCTIONAL PENETRATION TEST VALUES FOR THE LOT MET SPECIFICATION. NO ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO A DULL KNIFE WERE FOUND DURING THE DHR REVIEW AND THE PRODUCT WAS RELEASED ACCORDING TO THE MFR'S ACCEPTANCE CRITERIA. BECAUSE A SAMPLE WAS NOT RETURNED AND NO EVIDENCE OF NONCONFORMITY COULD BE FOUND IN THE LOT RECORD REVIEW, THE ROOT CAUSE FOR THE DULL KNIFE EXPERIENCED BY THE CUSTOMER CANNOT BE DETERMINED. SOME POTENTIAL CAUSES ARE REUSE, IMPROPER HANDLING, OR CONTACT WITH ANOTHER INSTRUMENT ON THE INSTRUMENT TRAY DURING PROCEDURE SETUP. ALL KNIVES ARE 100% INSPECTED BY TRAINED OPERATORS USING A MINIMUM OF 10X MAGNIFICATION DURING MANUFACTURING. ANY DEFECTS, SUCH AS DAMAGED TIPS AND CUTTING EDGES ARE REMOVED FROM THE LOT AND SCRAPPED. PENETRATION TESTING IS PERFORMED AND MONITORED DURING THE FINISHING PROCESS TO ENSURE THE SHARPNESS OF THE PRODUCT. (B)(4).
A CUSTOMER REPORTED THAT HE HAD DIFFICULTY MAKING THE INCISION FOR A CATARACT WITH INTRAOCULAR LENS IMPLANT PROCEDURE BECAUSE THE KNIFE DID NOT CUT. THIS ISSUE OCCURRED TWICE IN THE SAME OPERATING PROGRAM. THE KNIFE WAS EXCHANGED EACH TIME AND EACH CASE WAS COMPLETED WITH NO PT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 227469 | CUSTOM PAK | CONVENIENCE KIT | KYG | ALCON - BELGIUM/S.A. ALCON-COUVREUR N.V. | CUSTOM PAK | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CLEARCUT HP2 DUAL BEVEL ANGLED SLIT KNIFE 2.22MM |