FDA Adverse Event Malfunction Summary report: N

ACRYSOF

MDR report key: 3130329 · Received May 22, 2013

Report

Report Number
1119421-2013-00561
Event Type
Malfunction
Date Received
May 22, 2013
Date of Event
April 25, 2013
Report Date
April 25, 2013
Manufacturer
ALCON RESEARCH LTD/HUNTINGTON
Product Code
HQL
PMA / PMN Number
P930014
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT WAS NOT RETURNED FOR ANALYSIS. RESULTS FROM THE PRODUCT HISTORY RECORD REVIEW INDICATED THE PRODUCT MET RELEASE CRITERIA. THE CARTRIDGE PRODUCT HISTORY RECORD COULD NOT BE REVIEWED BECAUSE THE FACILITY DID NOT PROVIDE A LOT NUMBER OR ANY IDENTIFICATION TRACEABLE TO THE MANUFACTURING DOCUMENTATION. THE CUSTOMER INDICATED THE USE OF AN APPROVED CARTRIDGE, HAND PIECE AND VISCOELASTIC. THE PRODUCT INVESTIGATION COULD NOT IDENTIFY A ROOT CAUSE. THERE HAVE BEEN NO OTHER COMPLAINTS REPORTED IN THE LOT NUMBER. ATTEMPTS HAVE BEEN MADE TO OBTAIN ADDITIONAL INFORMATION. A COMPLETED QUESTIONNAIRE WAS RECEIVED ON (B)(4) 2013. (B)(4).

Description of Event or Problem · 1

A NURSE REPORTED A SCRATCHED INTRAOCULAR LENS (IOL) DURING A LENS IMPLANT PROCEDURE. THE LENS WAS INJECTED INTO THE CAPSULAR BAG AND TWO SCRATCH LINES IN A "V" SHAPE WERE SEEN IN THE CENTER OF THE OPTIC. THE SCRATCHED LENS WAS CUT, REMOVED AND REPLACED WITH THE SAME MODEL AND POWER LENS. A SECOND LENS WAS OPENED AND A SCRATCH WAS NOTED ON THAT LENS. THIS LENS WAS NOT USED ON THE PT. A THIRD LENS WAS OPENED AND USED TO COMPLETE THE PROCEDURE WITH NO PT IMPACT. A DELAY IN THE PROCEDURE WAS NOTED. ADDITIONAL INFORMATION WAS RECEIVED FROM THE NURSE WHO REPORTED A NORMAL PROCEDURE TAKES APPROX 30 MINUTES TO COMPLETE; HOWEVER, THIS PROCEDURE LASTED APPROX ONE AND ONE HALF HOURS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
227467 ACRYSOF INTRAOCULAR LENS HQL ALCON RESEARCH LTD/HUNTINGTON SN60WF 12158694

Patients

Seq Age Sex Outcome Treatment
1 72 YR DUOVISC| MONARC C CARTRIDGE| OZIL PHACO HANDPIECE| MONARCH II HANDPIECE