FDA Adverse Event Malfunction Summary report: N

CUSTOM PAK

MDR report key: 3130326 · Received May 22, 2013

Report

Report Number
1644019-2013-00058
Event Type
Malfunction
Date Received
May 22, 2013
Date of Event
April 24, 2013
Report Date
April 24, 2013
Manufacturer
ALCON - HOUSTON
Product Code
KYG
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ONE OPENED SAMPLE WAS RETURNED. EVAL OF THE SAMPLE SHOWED THE FOLLOWING RESULTS: THE SAMPLE WAS EXAMINED USING 10X MAGNIFICATION AND FOUND TO HAVE A DAMAGED TIP AND CUTTING EDGES. THE DEVICE HISTORY RECORDS (DHR) FOR THE TWO POSSIBLE LOTS WERE REVIEWED. FUNCTIONAL PENETRATION TEST VALUES FOR THIS LOT MET SPECIFICATION. A BENT TIP REJECTION WAS FOUND DURING THE DHR REVIEW. HOWEVER, THE SUSPECT PRODUCT WAS RE-INSPECTED AND THE PRODUCT WAS RELEASED ACCORDING TO THE MFR'S ACCEPTANCE CRITERIA. HOW OR WHEN THE BLADE BECAME DAMAGED CANNOT BE DETERMINED. THE DAMAGE TO THE RETURNED SAMPLE IS CONSISTENT WITH DAMAGE THAT CAN OCCUR WHEN THE BLADE CONTACTS ANOTHER SURFACE PRIOR TO USE. SOME POTENTIAL CAUSES COULD BE REUSE, IMPROPER HANDLING, OR CONTACT WITH ANOTHER INSTRUMENT ON THE INSTRUMENT TRAY DURING PROCEDURE SET UP. ALL KNIVES ARE 100% INSPECTED BY TRAINED OPERATORS USING A MINIMUM OF 10X MAGNIFICATION DURING MANUFACTURING. ANY DEFECTS, SUCH AS DAMAGED TIPS AND CUTTING EDGES, ARE REMOVED FROM THE LOT AND SCRAPPED. SHARPNESS TESTING IS PERFORMED AND MONITORED DURING THE FINISHING PROCESS TO ENSURE THE SHARPNESS OF THE PRODUCT. (B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT WHILE LOOKING THROUGH THE MICROSCOPE, HE NOTICED THAT THE KNIFE HAD A BURR ON THE TIP. THE CUSTOMER WAS ABLE TO BEND THE BURR FLAT USING FORCEPS. THE CUSTOMER HAD TO USE INCREASED PRESSURE TO MAKE THE INCISION DURING SURGERY. THE PROCEDURE WAS COMPLETED WITH NO HARM TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
227499 CUSTOM PAK CONVENIENCE KIT KYG ALCON - HOUSTON CUSTOM PAK 1378538H

Patients

Seq Age Sex Outcome Treatment
1 CLEARCUT SIDE PORT 1.2 MM ANGLED KNIFE