FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 3130257
·
Received May 22, 2013
Report
- Report Number
- 8020893-2013-01113
- Event Type
- Malfunction
- Date Received
- May 22, 2013
- Date of Event
- April 22, 2013
- Report Date
- April 22, 2013
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE 840 VENTILATOR HAD A BLANK SCREEN. THERE WAS NO PATIENT INVOLVEMENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE GRAPHICAL USER INTERFACE (GUI) CPU PCB AND POWER SUPPLY. THE UNIT PASSED EXTENDED SELF-TESTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 226456 | 840 VENTILATOR | VENTILATOR | CBK | COVIDIEN, FORMERLY NELLCOR | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |