LIGACLIP** ENDOSCOPIC CLIP APPLIER
Report
- Report Number
- 3005075853-2013-02584
- Event Type
- Malfunction
- Date Received
- May 24, 2013
- Date of Event
- April 26, 2013
- Report Date
- May 3, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). ADDITIONAL INFORMATION: WHICH FIRING OF THE DEVICE DID THIS EVENT OCCUR ON? ASKU. WHAT VESSEL OR STRUCTURE WAS THE DEVICE FIRED ON AT THE TIME OF THE EVENT? CYSTIC DUCT. WAS THE CLIP FULLY ADVANCED INTO THE JAWS PRIOR TO FIRING? ASKU. WAS THERE ANY TORQUING OR TWISTING OF THE DEVICE PRESENT AT THE TIME OF FIRING? ASKU. WAS ANY UNEXPECTED RESISTANCE FELT WHILE FIRING THE TRIGGER? ASKU. WERE ANY UNEXPECTED NOISES HEARD? ASKU. IF SO, WHEN? DID ANYTHING UNEXPECTED HAPPEN PRIOR TO THIS INCIDENT? NO. WAS THE DEVICE FIRED AFTER THIS INCIDENT IN OR OUT OF THE PATIENT? IN. WHAT WERE THE INDICATIONS FOR SURGERY? GALL BLADDER DISEASE. WHAT WAS FOUND? DOES THE PATIENT HAVE ANY RELEVANT HISTORY OF SURGICAL TREATMENTS? ASKU. IF SO, WHAT? DID SOMEBODY OTHER THAN THE PRIMARY SURGEON FIRE THE INSTRUMENT? NO. IF SO, WHOM? THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, RETAINED AND FORMED THE REMAINING CLIPS AS INTENDED. THE DEVICE WAS FOUND TO BE FULLY FUNCTIONAL AND CONFORMING TO OUR MANUFACTURING SPECIFICATIONS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE DEVICE WAS FIRED AND UPON OBSERVANCE OF THE CLIP IT HAD CUT THE CYSTIC DUCT. THEY USED ANOTHER LIKE DEVICE TO COMPLETE THE PROCEDURE. NO BLEEDING WAS REPORTED AT THIS TIME AND NO PATIENT CONSEQUENCE HAS BEEN REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 232043 | LIGACLIP** ENDOSCOPIC CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | K4CD2M |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |