MOLIFT SMART 150
Report
- Report Number
- 3004137175-2013-00007
- Event Type
- Malfunction
- Date Received
- May 21, 2013
- Date of Event
- November 8, 2011
- Report Date
- May 21, 2013
- Manufacturer
- ETAC AS (DIVISION ETAC SUPPLY GJOEVIK, NORWAY)
- Product Code
- FSA
- Removal / Correction Number
- 3004137175-2/15/13-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NO
- Reporter Occupation
- OTHER
Narratives
ETAC RECEIVED A TOTAL OF THREE REPORTS RELATED TO LOCKING CLAMPS OF THE LEG-SPREADING MECHANISM OF THE MOLIFT SMART 150 HOIST BETWEEN 2010 AND 2011. AS A RESULT OF FURTHER INVESTIGATION, ETAC IDENTIFIED A SINGLE BATCH OF LOCKING CLAMPS SUPPLIED BY A NEW, THIRD-PARTY SUPPLIER THAT WERE COMPOSED OF STEEL THAT WAS OF A HARDNESS GRADE BELOW ETAC'S HARDNESS SPECIFICATION FOR THE COMPONENT. BEGINNING IN 2011, ETAC UNDERTOOK A VOLUNTARY FIELD CORRECTION IN THE U.S. AND OTHER JURISDICTIONS TO REPLACE THE LOCKING CLAMPS OF ALL DELIVERED HOISTS THAT COULD HAVE INCLUDED THESE LOCKING CLAMPS. ETAC HAS REPORTED THE CORRECTION TO FDA (3004137175-2/15/13-001-C).
THE MOLIFT SMART 150 IS A FOLDABLE MOBILE HOIST AND IS INTENDED FOR LIFTING AND TRANSFERRING PTS. THE LEG-SPREADING MECHANISM OF THE HOIST FEATURES TWO LOCKING CLAMPS, LOCATED BEHIND A GRAY FOAM COVER ON EITHER SIDE OF THE CHASSIS. LOOSENING OF A LOCKING CLAMP OF THE LEG-SPREADING MECHANISM CAN RESULT IN ONE OF THE LEGS EXCEEDING THE NORMAL OPERATING AREA, LEADING TO DESTABILIZATION OF THE HOIST. ETAC RECEIVED A REPORT FROM (B)(6) THAT WHILE THE HOIST WAS BEING USED IN A HOME CARE SETTING, THE HOIST TOPPLED OVER WHILE A PERSON WAS BEING HOISTED. THERE WERE NO INJURIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 225699 | MOLIFT SMART 150 | LIFT, PATIENT, NON-AC-POWERED - FSA | FSA | ETAC AS (DIVISION ETAC SUPPLY GJOEVIK, NORWAY) | MOLIFE SMART 150 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |