FDA Adverse Event Malfunction Summary report: N

RPM RESPIRATORY GATING SYSTEM

MDR report key: 3130097 · Received May 21, 2013

Report

Report Number
2916710-2013-00011
Event Type
Malfunction
Date Received
May 21, 2013
Date of Event
April 24, 2013
Report Date
April 24, 2013
Manufacturer
VARIAN MEDICAL SYSTEMS, INC.
Product Code
IYE
PMA / PMN Number
K063270
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
PHYSICIST

Narratives

Additional Manufacturer Narrative · 1

ACTUAL DEVICE INVOLVED IN THE INCIDENT WAS EVALUATED. THOUGH STILL UNDER INVESTIGATION, VARIAN HAS DETERMINED THAT A MOR IS APPROPRIATE, AS THIS MALFUNCTION, SHOULD IT RECUR, COULD POTENTIALLY CAUSE A SERIOUS INJURY. ADDITIONAL FOLLOW-UP TO THIS MDR IS EXPECTED UPON COMPLETION OF THE INVESTIGATION. (B)(4).

Description of Event or Problem · 1

A PHYSICIST AT AN INSTITUTION THAT USES PHASE AND AMPLITUDE GATING IN THEIR SBRT PROGRAM HAS BEEN PERFORMING EXTENSIVE ANALYSIS OF THE DATA BEING GENERATED BY THE RPM RESPIRATORY GATING SYSTEM DURING TREATMENT OF LUNG PATIENTS WHO ARE BEING TREATED WITH RPM. THE RPM DATA IS EXPORTED AND THE PHYSICIST HAS USED THE EXPORTED DATA TO ANALYZE THE RPM BEHAVIOR STATING THAT THE RPM PHASE RESULTS DIFFER DRAMATICALLY DEPENDING UPON THE "LEARNING PHASE". THE PHYSICIST HAS MADE MANY ANALYSES OF THE RPM RESULTS, SHOWING THE DIFFERENCES USING AMPLITUDE VERSUS PHASE PLOTS. ACCORDING TO THE PHYSICIST, THE RPM RESULTS DIFFER DRAMATICALLY IN THE AMPLITUDE VERSUS PHASE PLOTS DEPENDING UPON THE "LEARNING PHASE". THE SAME PATIENT WILL HAVE DRAMATICALLY DIFFERENT AMPLITUDE VERSUS PHASE PLOTS IF THE "LEARNING PHASE" IS REPEATED. THE PHYSICIST HAS SEEN THIS SEVERAL PATIENTS AND IS ABLE TO REPRODUCE EXPECTED AND UNEXPECTED RESULTS MERELY BY "RELEARNING" THE BREATHING PATTERN, STATING THAT HE HAS TO PAY ATTENTION TO ENSURE THAT THE DESIRED RPM PHASE PATTERN IS OBSERVED. THE PHYSICIST NOTICED THIS BEHAVIOR ON THE FIRST SBRT PATIENT BEING TREATED AT THE INSTITUTE AND CORRECTED THE BEHAVIOR BEFORE CONTINUING WITH THE TREATMENT. NO INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
226185 RPM RESPIRATORY GATING SYSTEM ACCELERATOR, LINEAR, MEDICAL IYE VARIAN MEDICAL SYSTEMS, INC. H50

Patients

Seq Age Sex Outcome Treatment
1