FDA Adverse Event Malfunction Summary report: N

VASO VIEW HEMOPRO

MDR report key: 3130094 · Received May 21, 2013

Report

Report Number
2242352-2013-00473
Event Type
Malfunction
Date Received
May 21, 2013
Date of Event
April 22, 2013
Report Date
April 29, 2013
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
GEI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. IT SHOWED SIGNS OF CLINICAL USAGE AND EVIDENCE OF BLOOD. A VISUAL INSPECTION DETERMINED THAT THE THERE WAS A CRACK ON THE TIP OF THE COLD JAW. BASED UPON THE RECEIVED CONDITION OF THE DEICE, THE REPORTED COMPLAINT WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NON CONFORMANCE RECORDED IN THE LOT HISTORY. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING PREPARATION FOR AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE JAWS WERE COMING APART ON THE HEMOPRO. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
225698 VASO VIEW HEMOPRO ENDOSCOPIC VESSEL HARVESTING SYSTEM GEI MAQUET CARDIOVASCULAR, LLC VH-3000 25074131

Patients

Seq Age Sex Outcome Treatment
1 NA