FDA Adverse Event
Malfunction
Summary report: N
VASO VIEW HEMOPRO
MDR report key: 3130094
·
Received May 21, 2013
Report
- Report Number
- 2242352-2013-00473
- Event Type
- Malfunction
- Date Received
- May 21, 2013
- Date of Event
- April 22, 2013
- Report Date
- April 29, 2013
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- GEI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. IT SHOWED SIGNS OF CLINICAL USAGE AND EVIDENCE OF BLOOD. A VISUAL INSPECTION DETERMINED THAT THE THERE WAS A CRACK ON THE TIP OF THE COLD JAW. BASED UPON THE RECEIVED CONDITION OF THE DEICE, THE REPORTED COMPLAINT WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NON CONFORMANCE RECORDED IN THE LOT HISTORY. (B)(4).
Description of Event or Problem · 1
THE HOSPITAL REPORTED THAT DURING PREPARATION FOR AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE JAWS WERE COMING APART ON THE HEMOPRO. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.