MOLIFT SMART 150
Report
- Report Number
- 3004137175-2013-00006
- Event Type
- Malfunction
- Date Received
- May 21, 2013
- Date of Event
- May 3, 2011
- Report Date
- May 21, 2013
- Manufacturer
- ETAC AS (DIVISION ETAC SUPPLY GJOEVIK, NORWAY)
- Product Code
- FSA
- Removal / Correction Number
- 3004137175-2/15-13-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
ETAC RECEIVED A TOTAL OF THREE REPORTS RELATED TO LOCKING CLAMPS OF THE LEG-SPREADING MECHANISM OF THE MOLIFT SMART 150 HOIST BETWEEN 2010 AND 2011. AS A RESULT OF FURTHER INVESTIGATION, ETAC IDENTIFIED A SINGLE BATCH OF LOCKING CLAMPS SUPPLIED BY A NEW, THIRD-PARTY SUPPLIER THAT WERE COMPOSED OF STEEL THAT WAS OF A HARDNESS GRADE BELOW ETAC'S HARDNESS SPECIFICATION FOR THE COMPONENT. BEGINNING IN 2011, ETAC UNDERTOOK A VOLUNTARY FIELD CORRECTION IN THE U.S. AND OTHER JURISDICTIONS TO REPLACE THE LOCKING CLAMPS OF ALL DELIVERED HOISTS THAT COULD HAVE INCLUDED THESE LOCKING CLAMPS. ETAC HAS REPORTED THE CORRECTION TO FDA (3004137175-2-15-13-001-C).
THE MOLIFT SMART 150 IS A FOLDABLE MOBILE HOIST AND IS INTENDED FOR LIFTING AND TRANSFERRING PATIENTS. THE LEG-SPREADING MECHANISM OF THE HOIST FEATURES TWO LOCKING CLAMPS, LOCATED BEHIND A GRAY FOAM COVER ON EITHER SIDE OF THE CHASSIS. LOOSENING OF A LOCKING CLAMP OF THE LEG-SPREADING MECHANISM CAN RESULT IN ONE OF THE LEGS EXCEEDING THE NORMAL OPERATING AREA, LEADING TO DESTABILIZATION OF THE HOIST. A REPORT RECEIVED BY ETAC FROM THE (B)(6) STATED THAT A FATHER, WHEN PUTTING HIS SON TO BED, HOISTED THE SON TO MAXIMUM HEIGHT AND THAT WHILE THE FATHER WAS HOLDING BOTH HANDLES AND TURNING THE HOIST TOWARDS THE BED, THE HOIST BEGAN TO TOPPLE OVER. THE BOY LANDED ON THE BED AT AN ANGLE; THE HOIST DID NOT HIT HIM AND HE WAS UNINJURED. HOWEVER, AS THE HOIST BEGAN TO TIP, THE BOY'S BROTHER CAME FORWARD TO STOP THE HOIST AND WAS HIT ON THE SHINS BY THE HOIST AS IT FELL, CAUSING A GRAZE AN BRUISES TO BOTH SHINS. NO HOSPITAL, TREATMENT WAS GIVEN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 225182 | MOLIFT SMART 150 | LIFT, PATIENT, NON-AC-POWERED - FSA | FSA | ETAC AS (DIVISION ETAC SUPPLY GJOEVIK, NORWAY) | MOLIF SMART 150 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |