GREENLIGHT HPS BPH FIBER OPTIC
Report
- Report Number
- 2937094-2013-00601
- Event Type
- Malfunction
- Date Received
- May 20, 2013
- Date of Event
- February 13, 2013
- Report Date
- April 3, 2013
- Manufacturer
- AMERICAN MEDICAL SYSTEMS
- Product Code
- GEX
- PMA / PMN Number
- K062719
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- NURSE
Narratives
FIBER ANALYSIS: THE FIBER GLASS CAP WAS FOUND TO BE DETACHED; THE FIBER BROKEN PROXIMAL TO THE FIBER/CAP GLUE ZONE. THE FIBER CAP WAS NOT RETURNED BY THE CUSTOMER. THERE WAS NO INDICATION OR REPORT OF ANY PART OF THE DEVICE REMAINING INSIDE THE PATIENT. THE FIBER CONDITION COULD RESULT IN A FORWARD FIRING CONDITION OF THE SIDE-FIRING SURGICAL FIBER. THE MOST PROBABLE ROOT CAUSE OF THE DEVICE ISSUE WAS DETERMINED TO BE LOCALIZED HEAT ACCUMULATION/USER HANDLING, DUE TO ANATOMICAL/PROCEDURAL FACTORS ENCOUNTERED DURING THE PROCEDURE.
IT WAS REPORTED THAT WHILE USING THE SURGICAL FIBER DURING A PROSTATE PROCEDURE, THE FIBER TIP BURNT OUT (FIBER CEASED TO PROPERLY FUNCTION) AT 59,194 JOULES OF USE. THE CASE WAS COMPLETED USING A SECOND FIBER. THERE WAS NO INJURY REPORTED.