Narratives
Additional Manufacturer Narrative · 1
FIBER ANALYSIS: THE FIBERS GLASS CAP WAS FOUND TO BE DETACHED; THE FIBER BROKEN DISTAL TO THE FIBER/CAP GLUE ZONE. THE FIBER CAP WAS NOT RETURNED BY THE CUSTOMER. THERE WAS NO INDICATION OR REPORT OF ANY PART OF THE DEVICE REMAINING INSIDE THE PATIENT. THE FIBER CONDITION COULD RESULT IN A FORWARD FIRING CONDITION OF THE SIDE-FIRING SURGICAL FIBER. THE MOST PROBABLE ROOT CAUSE OF THE DEVICE ISSUE WAS DETERMINED TO BE LOCALIZED HEAT ACCUMULATION/USER HANDLING, DUE TO ANATOMICAL/PROCEDURAL FACTORS ENCOUNTERED DURING THE PROCEDURE.
Description of Event or Problem · 1
IT WAS REPORTED THAT WHILE USING THE SURGICAL FIBER DURING A PROSTATE PROCEDURE, THE FIBER TIP WAS DAMAGED AT 18,365 JOULES OF USE. THE CASE WAS COMPLETED USING A SECOND FIBER. THERE WAS NO INJURY REPORTED.