FDA Adverse Event Malfunction Summary report: N

GREENLIGHT HPS BPH FIBER OPTIC

MDR report key: 3130053 · Received May 20, 2013

Report

Report Number
2937094-2013-00598
Event Type
Malfunction
Date Received
May 20, 2013
Date of Event
March 16, 2013
Report Date
March 26, 2013
Manufacturer
AMERICAN MEDICAL SYSTEMS
Product Code
GEX
PMA / PMN Number
K062719
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

FIBER ANALYSIS: THE FIBERS GLASS CAP WAS FOUND TO BE DETACHED; THE FIBER BROKEN DISTAL TO THE FIBER/CAP GLUE ZONE. THE FIBER CAP WAS NOT RETURNED BY THE CUSTOMER. THERE WAS NO INDICATION OR REPORT OF ANY PART OF THE DEVICE REMAINING INSIDE THE PATIENT. THE FIBER CONDITION COULD RESULT IN A FORWARD FIRING CONDITION OF THE SIDE-FIRING SURGICAL FIBER. THE MOST PROBABLE ROOT CAUSE OF THE DEVICE ISSUE WAS DETERMINED TO BE LOCALIZED HEAT ACCUMULATION/USER HANDLING, DUE TO ANATOMICAL/PROCEDURAL FACTORS ENCOUNTERED DURING THE PROCEDURE.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE USING THE SURGICAL FIBER DURING A PROSTATE PROCEDURE, THE FIBER TIP WAS DAMAGED AT 18,365 JOULES OF USE. THE CASE WAS COMPLETED USING A SECOND FIBER. THERE WAS NO INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
223495 GREENLIGHT HPS BPH FIBER OPTIC POWERED SURGICAL LASER INSTRUMENT GEX AMERICAN MEDICAL SYSTEMS 10-2090 244H

Patients

Seq Age Sex Outcome Treatment
1