FDA Adverse Event
Other
Summary report: N
VISIONAIRE V
MDR report key: 3130041
·
Received May 22, 2013
Report
- Report Number
- 1319044-2013-00006
- Event Type
- Other
- Date Received
- May 22, 2013
- Date of Event
- April 25, 2013
- Report Date
- May 22, 2013
- Manufacturer
- AIRSEP CORP.
- Product Code
- CAW
- PMA / PMN Number
- K872534
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS FIRE WAS CAUSED BY EXTERNAL IGNITION OF THE CANNULA TUBING. PHOTOGRAPHS FROM THE FIRE SCENE SHOW THE CANNULA TUBING BURNT TO THE BED POST AND INTO THE CARPET. THE OXYGEN CONCENTRATOR CABINET WAS DAMAGED BY THE CANNULA BURNING BACK TO THE OXYGEN OUTLET WERE THE FIRE STOPPED. THE OXYGEN CONCENTRATOR IS OPERATIONAL.
Description of Event or Problem · 1
OXYGEN CONCENTRATOR INVOLVED IN A TRAILER FIRE. DAMAGE TO CARPET AND BED POST.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 227540 | VISIONAIRE V | OXYGEN CONCENTRATOR | CAW | AIRSEP CORP. | AS098-101 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR | Required Intervention |