FDA Adverse Event Malfunction Summary report: N

FREESTYLE FREEDOM LITE

MDR report key: 3130008 · Received May 24, 2013

Report

Report Number
2954323-2013-00284
Event Type
Malfunction
Date Received
May 24, 2013
Date of Event
April 28, 2013
Report Date
April 29, 2013
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED METER WAS RETURNED AND INVESTIGATED WITH RETAINED TEST STRIPS. THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION AND NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. ONE READING THE CUSTOMER REPORTED WAS FOUND IN THE METER MEMORY. GIVEN NO TEST STRIPS WERE RETURNED FOR INVESTIGATION, RETAINED TEST STRIP SAMPLES FROM THE SAME LOT REPORTED BY THE CUSTOMER (1350927) WERE TESTED WITH CONTROL SOLUTION. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION AND NO ERRORS WERE OBSERVED. THE COMPLAINT IS NOT CONFIRMED. (B)(4).

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 145 MG/DL AND 453 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
231479 FREESTYLE FREEDOM LITE BLOOD GLUCOSE MONITORING SYSTEM NBW 1350927

Patients

Seq Age Sex Outcome Treatment
1 58 YR